Feasibility and preliminary validation of a wearable sweat lactate sensor during incremental and steady-state exercise in trained runners

Francisco Javier Martínez-Noguera, Sofia Abud Zientek, Daniel Rojas, Xing Xuan, María Cuartero, Ramiro Alonso-Salinas et al. · Performance Enhancement & Health · 2026

Researchers tested a new wearable sensor that measures lactate in sweat as a non-invasive alternative to blood lactate testing during treadmill running. The sensor tracked lactate changes very accurately during incremental exercise tests, though it was less consistent during steady-state exercise due to individual differences in sweat production.
Takeaway: Consider sweat lactate sensors as an emerging option for monitoring training intensity during interval workouts, though blood lactate remains more reliable for steady-state exercise assessment.
Abstract (source)

Lactate measurement is a cornerstone in the assessment of metabolic stress during exercise. However, blood sampling is invasive and impractical in applied contexts. In this context, sweat has emerged as a promising biological matrix for continuous and non-invasive monitoring of biomarkers, although the validity and reliability of sweat lactate remain controversial. To evaluate the feasibility and preliminary reliability of a new portable sweat lactate sensor, comparing its performance with capillary blood lactate during an incremental protocol and a rectangular protocol in trained runners. Nine trained runners performed either incremental or constant load exercise tests on a treadmill. Lactate was measured simultaneously in capillary blood using the conventional lactate strip, and in sweat using a both a customized wearable sensor and ion chromatography as a reference

Method: . Kinetics, relative differences, correlations and linear regression models between

Methods: were analyzed. Sweat lactate was consistently higher than blood lactate in both protocols. During the incremental test, very high correlations were observed between blood lactate and sweat lactate measured with the sensor (r = 0.979; R² = 0.959; p < 0.0001), with parallel kinetics and a clear linear relationship. In the rectangular test, correlations were moderate to high (r = 0.539–0.988), although with greater interindividual variability. The differences between the sensor and ion chromatography were small compared to the blood–sweat differences. This pilot study demonstrates that the wearable sensor is capable of continuously and consistently detecting lactate dynamics in sweat, especially during incremental exercise. Although absolute quantification remains limited by sweat variability, noninvasive monitoring of lactate in sweat exhibit high potential as a complementary tool for monitoring metabolic stress during exercise.