Side differences in NIRS-derived muscle oxygen saturation during treadmill running
Lara Zwiener, Oliver Richrath, Philip Skotzke, Benjamin Volkamer-Kien, Sascha Schwindling, Tim Meyer · European Journal of Applied Physiology · 2026
Abstract
Purpose: Near-infrared spectroscopy (NIRS) offers a promising non-invasive
Method: to assess muscle oxygenation in real time. While asymmetries in muscle oxygen saturation (SmO 2 ) between dominant and non-dominant legs have been studied in cycling, data in running are limited. Such differences may affect NIRS use in training control. This study investigated SmO 2 differences between dominant and non-dominant vastus lateralis during treadmill running at three workloads using a wearable continuous-wave NIRS device.
Methods: Forty-eight participants (20 female, Age = 21.6 ± 1.5 years) performed two graded treadmill exercise tests, 14 days apart. The Moxy Monitor was placed bilaterally on the vastus lateralis to measure SmO 2 and analysed across three individualized workloads (LOW, MID, HIGH). Linear mixed models were used with participant as random, and side and workload as fixed effects. To assess adipose tissue thickness (ATT) effects, a 7mm cut-off defined two subgroups (ATT 7mm). Bland-Altman plots examined mean bias and limits of agreement.
Results: No significant side effects were found for either group (ALL p = .759, ATT 7mm p = .890). Bland-Altman analysis revealed a small mean bias (ALL: -0.34%, ATT 7mm: -0.22%) with wide limits of agreement between the legs.
Conclusion: No systematic differences between SmO 2 response between the legs were found. However, wide limits of agreement suggest that individual side differences may exist. A single NIRS device may sufficiently reflect bilateral SmO 2 responses at the group level, while individual side-to-side-comparisons should be interpreted with caution.